1967 J. Sci. Instrum. 44 360-362 doi: 10.1088/0950-7671/44/5/308
![]()
|
||||
Abstract. An automatic pressure stepping control and re-cycling system for use with a Fabry-Pérot interferometer is described. The latter uses narrow band interference filters to isolate the spectral region of interest (Hβ at λ = 4861 Å) and is scanned over wavelength ranges from 6 to 24 Å units centred on the above wavelength. A pre-selected programme of spectral elements can be sampled with an accuracy to ±1 torr and this corresponds to ±0.002 Å, when nitrogen is used as the control gas. The interferometer is designed to operate at resolutions from 0.25 to 1.0 Å at a finesse of 40 and with corresponding plate separations between 125 and 40 μm.
Print publication: Issue 5 (May 1967)| Post to CiteUlike | | Post to Connotea | | Post to Bibsonomy |
|
Journals Home | Journals List | EJs Extra | This Journal | Search | Authors | Referees | Librarians | User Options | Help | Recommend this journal EndNote, ProCite ® and Reference Manager ® are registered trademarks of ISI Researchsoft. Copyright © Institute of Physics and IOP Publishing Limited 2009. Use of this service is subject to compliance with the terms and conditions of use. In particular, reselling and systematic downloading of files is prohibited. Help: Cookies | Data Protection. |
|
| |