Measurements of wavelength shifting (WLS) fibre readout for a highly multiplexed, depth-encoding PET detector

Huini Du et al 2007 Phys. Med. Biol. 52 2499-2514   doi: 10.1088/0031-9155/52/9/011  Help

   PDF (374 KB) | References | Articles citing this article

Huini Du, Yongfeng Yang and Simon R Cherry
Department of Biomedical Engineering, University of California-Davis, One Shields Avenue, CA 95616, USA
E-mail: hdu@ucdavis.edu

Abstract. We are attempting to develop a laboratory PET scanner for mouse imaging that utilizes far fewer detectors and channels of electronics, thus reducing cost, whilst retaining state-of-the-art performance. The detectors are based on LSO arrays read out by wavelength shifting (WLS) fibres placed on the top and the bottom of the arrays. Depth of interaction information will be obtained from the ratio of the signals at either end of the array. For acceptable performance, it is critical to maximize collection of light photons from the ends of the fibres and to minimize the optical crosstalk between adjacent fibres. Factors which can affect the light collection and crosstalk were studied, including coupling materials between fibres and crystals, reflectors wrapped around the fibre sides and ends, fibre size and shape, and number of layers of fibre cladding. Properties of WLS fibres such as the transmission attenuation and transverse absorption were also studied. The light yield from 2 × 2 × 10 mm3 LSO crystals collected from the end of a 2 × 2 × 30 mm3 WLS fibre was up to 24% (typical values 16–20%) of that obtained by direct coupling of the LSO crystal. This light collection efficiency appears to be sufficient for decoding interaction position in these detectors.

Print publication: Issue 9 (7 May 2007)
Received 8 January 2007, in final form 27 February 2007
Published 11 April 2007

Bookmark and Share Post to CiteUlike | Post to Connotea | Post to Bibsonomy

 


Find related articles






Article options

Authors & Referees

 
Content finder
  Full Search
  Help


  
Setup information is available for Adobe Acrobat.
EndNote, ProCite ® and Reference Manager ® are registered trademarks of ISI Researchsoft.
Copyright © Institute of Physics and IOP Publishing Limited 2009.
Use of this service is subject to compliance with the Terms and Conditions of use. In particular, reselling and systematic downloading of files is prohibited.
Help: Cookies | Data Protection. Privacy policy Disclaimer