Optimization of megavoltage CT scan registration settings for brain cancer treatments on tomotherapy

Curtis Woodford et al 2007 Phys. Med. Biol. 52 N185-N193   doi: 10.1088/0031-9155/52/8/N04  Help

   PDF (248 KB) | References | Articles citing this article

Curtis Woodford, Slav Yartsev and Jake Van Dyk
London Regional Cancer Program, London Health Sciences Centre, 790 Commissioners Road East, London, Ontario, Canada
E-mail: slav.yartsev@lhsc.on.ca

Abstract. This study aims to determine the settings that provide the optimal clinical accuracy and consistency for the registration of megavoltage CT (MVCT) with planning kilovoltage CT image sets on the Hi-ART tomotherapy system. The systematic offset between the MVCT and the planning kVCT was determined by registration of multiple MVCT scans of a head phantom aligned with the planning isocentre. Residual error vector lengths and components were used to quantify the alignment quality for the phantom shifted by 5 mm in different directions obtained by all 27 possible combinations of MVCT inter-slice spacing, registration techniques and resolution. MVCT scans with normal slices are superior to coarse slices for registration of shifts in the superior–inferior, lateral and anterior–posterior directions. Decreasing the scan length has no detrimental effect on registration accuracy as long as the scan lengths are larger than 24 mm. In the case of bone technique and fine resolution, normal and fine MVCT scan slice spacing options give similar accuracy, so normal mode is preferable due to shorter procedure and less delivered dose required for patient set-up. A superior–inferior field length of 24–30 mm, normal slice spacing, bone technique, and fine resolution is the optimum set of registration settings for MVCT scans of a Rando head phantom acquired with the Hi-ART tomotherapy system, provided the registration shifts are less than 5 mm.

Print publication: Issue 8 (21 April 2007)
Received 7 September 2006, in final form 9 February 2007
Published 27 March 2007

Bookmark and Share Post to CiteUlike | Post to Connotea | Post to Bibsonomy

 


Find related articles






Article options

Authors & Referees

 
Content finder
  Full Search
  Help


  
Setup information is available for Adobe Acrobat.
EndNote, ProCite ® and Reference Manager ® are registered trademarks of ISI Researchsoft.
Copyright © Institute of Physics and IOP Publishing Limited 2009.
Use of this service is subject to compliance with the Terms and Conditions of use. In particular, reselling and systematic downloading of files is prohibited.
Help: Cookies | Data Protection. Privacy policy Disclaimer