A numerical study of transcranial focused ultrasound beam propagation at low frequency

Xiangtao Yin et al 2005 Phys. Med. Biol. 50 1821-1836   doi: 10.1088/0031-9155/50/8/013  Help

   PDF (1021 KB) | References | Articles citing this article

Xiangtao Yin and Kullervo Hynynen
Department of Radiology, Brigham and Women's Hospital, Harvard Medical School, 221 Longwood Ave, Boston, MA 02115, USA
E-mail: xyin@bwh.harvard.edu

Abstract. The feasibility of transcranial ultrasound focusing with a non-moving phased array and without skull-specific aberration correction was investigated using computer simulations. Three cadaver skull CT image data sets were incorporated into an acoustic wave transmission model to simulate transskull ultrasound wave propagation. Using a 0.25 MHz hemispherical array (125 mm radius of curvature, 250 mm diameter, 24 255 elements), the simulated beams could be focused and steered with transducer element driving phases and amplitude adjusted for focal beam steering in water (water-path). A total of 82 foci, spanning wide ranges of distance in the three orthogonal dimensions, were simulated to test the focal beam steering capability inside the three skulls. The acoustic pressure distribution in a volume of 20 × 20 × 20 mm3 centred at each focus was calculated with a 0.5 mm spacing in each axis. Clearly defined foci were retained through the skulls (skull-path) in most cases. The skull-path foci were on average 1.6 ± 0.8 mm shifted from their intended locations. The −3 dB skull-path beam width and length were on average 4.3 ± 1.0 mm and 7.7 ± 1.8 mm, respectively. The skull-path sidelobe levels ranged from 25% to 55% of the peak pressure values. The skull-path peak pressure levels were about 10%–40% of their water-path counterparts. Focusing low-frequency beam through skull without skull-specific aberration correction is possible. This method may be useful for applying ultrasound to disrupt the blood–brain barrier for targeted delivery of therapeutic or diagnostic agents, or to induce microbubbles, or for other uses of ultrasound in brain where the required power levels are low and the sharp focusing is not needed.

Print publication: Issue 8 (21 April 2005)
Received 8 November 2004, in final form 2 February 2005
Published 6 April 2005

Bookmark and Share Post to CiteUlike | Post to Connotea | Post to Bibsonomy

 


Find related articles






Article options

Authors & Referees

 
Content finder
  Full Search
  Help


  
Setup information is available for Adobe Acrobat.
EndNote, ProCite ® and Reference Manager ® are registered trademarks of ISI Researchsoft.
Copyright © Institute of Physics and IOP Publishing Limited 2009.
Use of this service is subject to compliance with the terms and conditions of use. In particular, reselling and systematic downloading of files is prohibited.
Help: Cookies | Data Protection.