1993 Phys. Med. Biol. 38 1937-1955 doi: 10.1088/0031-9155/38/12/016
![]()
|
||||
Abstract. An indirect comparison of the absorbed dose to water standards of the PTB and the NRC was carried out for 18/20 MV X-rays using five ionization chambers as transfer instruments. The absorbed dose standard of the PTB is based on the total absorption of 5.6 MeV electrons in Fricke solution. The NRC standard uses Fricke solution whose calibration is based on measurements made with a water calorimeter and a calculation of the heat defect. For high-energy X-rays, the difference found between the standards of 0.4% is within the expected uncertainties. The comparison was linked to existing 60Co absorbed dose to water and air kerma standards by comparing measurements of those quantities at the PTB, NRC and BIPM. Agreement was better than 0.7% in all cases. For the ionization chambers used in this study, the absorbed dose calibration factors for 18/20 MV X-rays are about 2% lower than those for 60Co gamma-rays.
Print publication: Issue 12 (December 1993)| Post to CiteUlike | | Post to Connotea | | Post to Bibsonomy |
|
Journals Home | Journals List | EJs Extra | This Journal | Search | Authors | Referees | Librarians | User Options | Help | Recommend this journal EndNote, ProCite ® and Reference Manager ® are registered trademarks of ISI Researchsoft. Copyright © Institute of Physics and IOP Publishing Limited 2009. Use of this service is subject to compliance with the Terms and Conditions of use. In particular, reselling and systematic downloading of files is prohibited. Help: Cookies | Data Protection. Privacy policy Disclaimer |