SUPPLEMENTARY COMPARISON

Final report on supplementary comparison APMP.M.P-S2 (bilateral comparison) at a nominal pressure of 0.05 Pa

Pardeep Mohan et al 2008 Metrologia 45 07003   doi: 10.1088/0026-1394/45/1A/07003  Help


Pardeep Mohan1 and Patrick J Abbott2
1 National Physical Laboratory, Dr K S Krishnan Marg, New Delhi-110012, India
2 National Institute of Standards and Technology, 100 Bureau Drive, Gaithersburg, MD 20899, USA

Abstract. A supplementary bilateral comparison was carried out to compare two different pressure standards at a nominal pressure of 0.05 Pa. The two standards used different principles for generating the nominal pressure; one based on the static expansion method (NPL, India), and the other based on the orifice flow or dynamic expansion method (NIST, USA). The transfer standard consisted of a pair of spinning rotor gauges whose accommodation coefficients were measured at a nominal pressure of 0.05 Pa at the two laboratories. The nominal pressure value for the comparison was chosen because (i) it can be easily generated by the respective standards with high accuracy and (ii) it had not been covered in the earlier two key comparisons, namely, CCM.P-K3 and CCM.P-K4, which the two NMIs had participated in. NPLI served as the pilot lab, and provided the stainless steel rotors; each lab used its own thimbles and electronics. The rotors were hand-carried from NPLI to NIST at atmospheric pressure in small, insulation-packed glass vials to prevent the rotors from moving during transit; otherwise no special transit precautions were taken.

The pressure value of 0.05 Pa generated by the NPLI and NIST showed good agreement within their combined uncertainties, proving thereby that the two standards were equivalent to each other for the pressure of 0.05 Pa.

Main text. To reach the main text of this Paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.

The final report has been peer-reviewed and approved for publication by the APMP, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).

Bookmark and Share Post to CiteUlike | Post to Connotea | Post to Bibsonomy

 

Find related articles





Article options

Authors & Referees

BEC Matters!author services
 
Content finder
  Full Search
  Help


  
EndNote, ProCite ® and Reference Manager ® are registered trademarks of ISI Researchsoft.
Copyright © Institute of Physics and IOP Publishing Limited 2009.
Use of this service is subject to compliance with the terms and conditions of use. In particular, reselling and systematic downloading of files is prohibited.
Help: Cookies | Data Protection.