KEY COMPARISON

APMP/TCRI key comparison report of measurement of air kerma for medium-energy x-rays (APMP.RI(I)-K3)

J H Lee et al 2008 Metrologia 45 06012   doi: 10.1088/0026-1394/45/1A/06012  Help


J H Lee1, W S Hwang1, L H Kotler2, D V Webb2, L Büermann3, D T Burns4, M Takeyeddin5, V V Shaha6, S Srimanoroth7, A Meghzifene8, S H Hah9, K J Chun9, T B Kadni10, N Takata11 and Z Msimang12
1 Institute of Nuclear Energy Research, Longtan, Taiwan
2 Australian Radiation Protection and Nuclear Safety Agency, Yallambie, Australia
3 Physikalisch Technische Bundesanstalt, Braunschweig, Germany
4 Bureau International de Poids et Mesures, Sèvres, France
5 Atomic Energy Commission, Damascus, Syria
6 Bhabha Atomic Research Centre, Mumbai, India
7 Department of Medical Sciences, Nonthaburi, Thailand
8 International Atomic Energy Agency, Vienna, Austria
9 Korea Research Institute of Standards and Science, Yusong, Korea
10 Malaysian Nuclear Agency, Kajang, Malaysia
11 National Metrology Institute of Japan, Tsukuba, Japan
12 National Metrology Institute of South Africa, Pretoria, South Africa

Abstract. The APMP/TCRI Dosimetry Working Group performed the APMP.RI(I)-K3 key comparison of measurement of air kerma for medium-energy x-rays (100 kV to 250 kV) between 2000 and 2003. In total, 11 institutes took part in the comparison, among which 8 were APMP member laboratories. Two commercial cavity ionization chambers were used as transfer instruments and circulated among the participants. All the participants established the 100 kV, 135 kV, 180 kV and 250 kV x-ray beam qualities equivalent to those of the BIPM. The results showed that the maximum difference between the participants and the BIPM in the medium-energy x ray range, evaluated using the comparison data of the linking laboratories ARPANSA and PTB, is less than 1.4%. The degrees of equivalence between the participants are presented and this comparison confirms the calibration capabilities of the participating laboratories.

Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.

The final report has been peer-reviewed and approved for publication by the CCRI Section I, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).

Bookmark and Share Post to CiteUlike | Post to Connotea | Post to Bibsonomy

 

Find related articles





Article options

Authors & Referees

IOP Journal Archiveeprintweb.org - Your address for E prints
 
Content finder
  Full Search
  Help


  
EndNote, ProCite ® and Reference Manager ® are registered trademarks of ISI Researchsoft.
Copyright © Institute of Physics and IOP Publishing Limited 2009.
Use of this service is subject to compliance with the Terms and Conditions of use. In particular, reselling and systematic downloading of files is prohibited.
Help: Cookies | Data Protection. Privacy policy Disclaimer