| Journals Home | Journals List | EJs Extra | This Journal | Search | Authors | Referees | Librarians | User Options | Help | | ||||
SHORT COMMUNICATION
2007 Metrologia 44 L53-L56 doi: 10.1088/0026-1394/44/6/N02
![]()
|
||||
Abstract. A re-evaluation of the BIPM standard for air kerma in 60Co radiation has been made. The changes to the air-kerma rate determination arise from four sources: (i) the results of Monte Carlo calculations of correction factors for the standard; (ii) a re-evaluation of the correction factor for saturation; (iii) a new evaluation of the air volume of the standard using an experimental chamber of variable volume; (iv) the adoption of a new reference beam at the BIPM. The combined effect of these changes is an increase in the BIPM determination of air-kerma rate by the factor 1.0054 and a reduction of the relative standard uncertainty of this determination to 1.5 parts in 103. A full uncertainty budget is presented. The new standard is effective from 1 November 2007.
Print publication: Issue 6 (December 2007)| Post to CiteUlike | | Post to Connotea | | Post to Bibsonomy |
|
Journals Home | Journals List | EJs Extra | This Journal | Search | Authors | Referees | Librarians | User Options | Help | Recommend this journal EndNote, ProCite ® and Reference Manager ® are registered trademarks of ISI Researchsoft. Copyright © Institute of Physics and IOP Publishing Limited 2009. Use of this service is subject to compliance with the Terms and Conditions of use. In particular, reselling and systematic downloading of files is prohibited. Help: Cookies | Data Protection. Privacy policy Disclaimer |