Abstract. Sources of bias in the National Institute of Standards and Technology ozone Standard Reference Photometer (SRP) maintained by the Bureau International des Poids et Mesures have been investigated. A relative bias of −0.4% in the ozone mole fraction measurement caused by a temperature gradient in the gas cells of the instrument was characterized and corrected for in a modified version of the instrument. A second relative bias of +0.5% due to the multiple reflections of light within the gas cells was also corrected. The Guide to the Expression of Uncertainty in Measurement approach was used to develop an uncertainty budget for the modified SRP, including a relative value for the ozone absorption cross-section uncertainty of 2.1% (k = 2). The measurement uncertainty for the bias-corrected SRPs is enlarged compared with earlier studies, but their comparability improved.
Print publication: Issue 5 (October 2006)| Post to CiteUlike | | Post to Connotea | | Post to Bibsonomy |
|
Journals Home | Journals List | EJs Extra | This Journal | Search | Authors | Referees | Librarians | User Options | Help | Recommend this journal EndNote, ProCite ® and Reference Manager ® are registered trademarks of ISI Researchsoft. Copyright © Institute of Physics and IOP Publishing Limited 2009. Use of this service is subject to compliance with the Terms and Conditions of use. In particular, reselling and systematic downloading of files is prohibited. Help: Cookies | Data Protection. Privacy policy Disclaimer |