Abstract. A method is described to determine the volume of a mass standard from weighings made in air of differing density against a reference volume artefact. Standard uncertainties in volume of 2 parts in 105 (at 1 kg) or better are readily achievable with this method, using adequate measuring facilities. The advantages of the method are that volume determination and mass calibration may be carried out simultaneously and that the mass standard is subjected to a variation in air density of only ±5 % or less, compared with conventional hydrostatic weighing methods where the mass standard is immersed in water. Examples are presented of the use of the method for confirming values assigned to mass standards and for full volume determination, together with uncertainty calculations.
Print publication: Issue 1 (February 2001)| Post to CiteUlike | | Post to Connotea | | Post to Bibsonomy |
|
Journals Home | Journals List | EJs Extra | This Journal | Search | Authors | Referees | Librarians | User Options | Help | Recommend this journal EndNote, ProCite ® and Reference Manager ® are registered trademarks of ISI Researchsoft. Copyright © Institute of Physics and IOP Publishing Limited 2009. Use of this service is subject to compliance with the terms and conditions of use. In particular, reselling and systematic downloading of files is prohibited. Help: Cookies | Data Protection. |