1996 Metrologia 33 197-206 doi: 10.1088/0026-1394/33/3/2
![]()
|
||||
Abstract. The primary purpose of the comparison was to establish the equivalence of the calibration of infrared radiation thermometers within Europe in the temperature range 800 °C to 2000 °C. Seven EU national metrological institutes (NMIs) participated. The equivalence of calibrations was determined by circulating a transfer standard infrared radiation thermometer between the participating laboratories. The calibrations obtained were compared with the reference calibration performed by the co-ordinating laboratory (National Physical Laboratory). It was found that, after correction for the size-of-source effect, the calibrations of the participating laboratories nearly all agree to better than 0,1 % of temperature.
Print publication: Issue 3 (1996)| Post to CiteUlike | | Post to Connotea | | Post to Bibsonomy |
|
Journals Home | Journals List | EJs Extra | This Journal | Search | Authors | Referees | Librarians | User Options | Help | Recommend this journal EndNote, ProCite ® and Reference Manager ® are registered trademarks of ISI Researchsoft. Copyright © Institute of Physics and IOP Publishing Limited 2009. Use of this service is subject to compliance with the Terms and Conditions of use. In particular, reselling and systematic downloading of files is prohibited. Help: Cookies | Data Protection. Privacy policy Disclaimer |