| Journals Home | Journals List | EJs Extra | This Journal | Search | Authors | Referees | Librarians | User Options | Help | | ||||
1983 Metrologia 19 37-47 doi: 10.1088/0026-1394/19/1/007
![]()
|
||||
Abstract. Seventeen standard platinum resistance thermometers from five commercial sources have been subjected to repeated thermal cycling over all or most of the range -196 °C to +650 °C to determine whether strain effects occur in the platinum sensors, and to what precision they affect temperature measurements. It is found that the resistance at the triple point of water, when measured repeatedly during thermal cycling, does undergo reversible changes in many thermometers that can be attributed to either elastic or anelastic strain. The magnitude of this resistance irreproducibility is usually between the equivalent of 0.1 m °C and 1.8 m °C near 0 °C. It is concluded that thermal strain effects pose a small, but significant, limitation on the ultimate performance of most of these thermometers.
Print publication: Issue 1 (1983)| Post to CiteUlike | | Post to Connotea | | Post to Bibsonomy |
|
Journals Home | Journals List | EJs Extra | This Journal | Search | Authors | Referees | Librarians | User Options | Help | Recommend this journal EndNote, ProCite ® and Reference Manager ® are registered trademarks of ISI Researchsoft. Copyright © Institute of Physics and IOP Publishing Limited 2009. Use of this service is subject to compliance with the Terms and Conditions of use. In particular, reselling and systematic downloading of files is prohibited. Help: Cookies | Data Protection. Privacy policy Disclaimer |