| Journals Home | Journals List | EJs Extra | This Journal | Search | Authors | Referees | Librarians | User Options | Help | | ||||
1979 J. Phys. E: Sci. Instrum. 12 216-224 doi: 10.1088/0022-3735/12/3/018
![]()
|
||||
Abstract. The development of pyrometry with time resolution in the microsecond range is required by high-speed thermophysical property experiments with millisecond duration. The general design criteria of a high-speed pyrometer with such a resolution are discussed and the contribution of several parameters is analysed to optimise the pyrometer performance with respect to the temperature-time characteristic of the event to be measured. Salient features of a pyrometer built on the basis of these criteria are described, with the details of the procedures necessary for its calibration. Monochromatic operation either near 0.65 mu m or near 1 mu m and a response time to a step input of 5 mu s make the instrument adequate for measuring temperature with a time resolution in the microsecond range. Emphasis is placed on the inaccuracy of temperature measurements made with the instrument.
Print publication: Issue 3 (March 1979)| Post to CiteUlike | | Post to Connotea | | Post to Bibsonomy |
|
Journals Home | Journals List | EJs Extra | This Journal | Search | Authors | Referees | Librarians | User Options | Help | Recommend this journal EndNote, ProCite ® and Reference Manager ® are registered trademarks of ISI Researchsoft. Copyright © Institute of Physics and IOP Publishing Limited 2010. Use of this service is subject to compliance with the Terms and Conditions of use. In particular, reselling and systematic downloading of files is prohibited. Help: Cookies | Data Protection. Privacy policy Disclaimer |